REVIEW
Department of Cell Biology, Yale University School of Medicine, New Haven, Connecticut lukesjulson{at}gmail.com
| Abstract |
|---|
| Introduction |
|---|
|
|
|---|
The choice of optimal imaging strategies for a given biological application thus requires a careful balance of tradeoffs. This is particularly true for the detection of all-or-none events, such as action potentials (APs) or vesicle fusion events (6, 26), in noisy single-trial data, where across-trial averaging cannot be used. The analysis of such experiments, as well as the design of optical reporters and instrumentation, requires an understanding of the basic principles of signal detection. This review illustrates these principles, using as an example the detection of APs with optical voltage reporters.
| Optical Voltage Reporters |
|---|
|
|
|---|
Since the electric field is localized to the plasma membrane, voltage reporters require a probe that sits in the plasma membrane itself. The change in membrane potential during an AP affects the probe molecule in a way that alters an optically observable parameter, such as fluorescence. All published voltage reporters with fast kinetics fall into four classes: fast voltage-sensitive dyes (VSDs), two-component VSDs, hybrid GFP-dye systems, and fusions of GFP with ion channels.
Fast VSDs (FIGURE 1A
) were the first voltage reporters developed by Cohen and colleagues (12, 40). VSDs partition into one leaflet of the plasma membrane, where they are influenced by the membrane electric field. The three major mechanisms of fluorescence modulation by membrane voltage are believed to be potential-dependent spectral shifts (electrochromism), dipole reorientation, and alterations of monomer-dimer equilibria (15, 17, 25, 47).
|
Hybrid GFP-dye systems (FIGURE 1C
) are analogous to the two-component VSD systems, except that the stationary dye is replaced by a GFP anchored to the cytoplasmic face of the plasma membrane. This design was also pioneered by the Tsien group (45). Recently, the approach has been extended in the hybrid voltage sensor (hVOS) system (10), in which a nonfluorescent mobile dye quenches the fluorescence of the membrane-bound GFP.
Voltage reporters based on GFP-channel fusions (FIGURE 1D
) were introduced by Siegel and Isacoff with the development of FlaSh, or "Fluorescent Shaker," a fusion of GFP with the Shaker potassium channel (43). Voltage-gated cation channels undergo a series of conformational changes with membrane voltage and in fact underlie the AP itself. Although the mechanism of fluorescence modulation is not clear, GFP-channel fusions operate according to the principle that normal conformational changes of the channel are translated to changes in GFP fluorescence. Several GFP-channel fusions exist: FlaSh, VSFP1 (38), and SPARC (3), as well as several modified versions of FlaSh (21).
| Wide-Field vs. Laser Scanning Microscopy |
|---|
|
|
|---|
Biosynthetic voltage reporters potentially solve the problem of indiscriminate labeling (see below), whereas laser scanning microscopy solves the problem of detecting out-of-focus fluorescence by optical sectioning, or selectively imaging a single plane in the specimen. There are three common forms of laser scanning microscopy with optical sectioning capability: confocal, two-photon (14), and second harmonic generation (SHG) microscopy (8). In confocal microscopy, a pinhole rejects out-of-focal plane light, whereas in two-photon and SHG microscopy, emission light is generated only at the point of focus.
Despite superior depth penetration and resolution in thick tissue, laser scanning microscopy introduces two serious technical difficulties for recording short-duration optical events, such as APs. First, because the beam must be scanned point by point over the specimen, image acquisition is much slower than wide-field imaging: frame rates for typical laser scanning microscopes are in the range of 14 Hz, which precludes imaging of optical events of ~1 ms in duration. In contrast, wide-field imaging with high-speed CCD cameras or photodiode arrays can acquire images at rates >1 kHz (5). High-speed multiphoton scanning microscopes based on resonant galvanometers (16, 49) or acousto-optic deflectors (24, 36) can also scan individual lines (but not full frames) at speeds in the range of 716 kHz. However, line scanning samples only a tiny fraction of the field, removing much of the advantage of imaging over electrodes. The development of microscopes with random beam addressability in two dimensions, which are capable of scanning tens of locations at >1 kHz (22, 39), partially overcomes this limitation. It has also been demonstrated that it is possible to perform optical sectioning under special wide-field multiphoton illumination conditions (31). Nevertheless, scanning speed is a significant limitation for the vast majority of existing laser scanning microscopes.
The second, more fundamental, drawback of scanning microscopy is that the total number of photons detected in a given time interval is small relative to wide-field imaging. For example, Baker et al. (5) observe a 1,000-fold decrease in the number of photons collected using two-photon microscopy relative to wide-field microscopy under similar conditions. This is for two reasons. First, the excitation beam illuminates each point in the image only once per frame, whereas in wide-field imaging each point is illuminated continuously. Additionally, wide-field imaging collects light from an entire cell, not just a single plane through the cell. The effect of low photon flux on AP detection can best be quantified from the standpoint of signal detection theory.
| Signal Detection Theory |
|---|
|
|
|---|
|
![]() |
where
is the mean number of photons in each bin. The right side invokes the central limit theorem, which states that when
is large, the Poisson distribution can be approximated as a Gaussian, or normal, distribution with a mean of
and a standard deviation of 
. For experimentally realistic values of
, the Gaussian approximation is accurate, and we therefore adopt it for convenience.
To illustrate the effect of shot noise on signal detection, consider the simple case where an AP causes a voltage reporter to increase fluorescence by 10% for 1 ms. If each 1-ms bin contains 1,000 photons on average when no AP occurs, bins containing APs will contain 1,100 photons on average. Revisiting our fictitious zero-noise example (FIGURE 2A
), we find it is easy to set a threshold that distinguishes an AP with perfect accuracy. Taking shot noise into consideration (FIGURE 2B
), we see that the "noise" and "signal + noise" distributions overlap. As a consequence, it is no longer possible to achieve perfect detection efficiency with no false positives: an acceptable tradeoff must be determined. Since we do not know in advance how many APs to expect, the simplest criterion to guide us is the Neyman-Pearson decision rule (48). This decision rule sets a threshold to distinguish between bins containing APs and those not containing APs. If the number of photons in a bin is greater than this threshold, we decide an AP has occurred (D1). The threshold is set arbitrarily to the level of false positives deemed acceptable, with increasing false positives leading to increased detection probability. For example, if we are willing to tolerate 10 spurious APs per second, a large number, the "false-positive" probability (PFP) is 102 (10 out of the 1,000 1-ms time bins/s will contain a spurious AP). From the "noise" distribution p(x | H0), the threshold t is determined such that the area under the part of the curve falling above the threshold is equal to 102 (red region, FIGURE 2B
). In this case,
![]() |
where
is the standard normal cumulative distribution function. If more than 1,074 photons were detected, we decide an AP has occurred. Working with this threshold, the probability of successful AP detection (PD) can be calculated from the "signal + noise" distribution p(x | H1) as the area under the curve to the right of the threshold (green region, FIGURE 2B
):
![]() |
In this case, shot noise limits detection to only 78% of APs, despite the high false-positive rate of 10 APs/s. If the cell were firing at 10 Hz, ~8 (out of 10) real APs/s would be detected along with 10 spurious ones. In other words, <45% of the APs detected would correspond to real APs, and we would overestimate the neurons true firing rate of 10 Hz by almost twofold as 18 Hz. These significant errors are a direct and unavoidable consequence of a low photon count; if the photon count increases, the situation improves dramatically, as discussed below.
There are other experimental noise sources in addition to shot noise, such as photodetector excess noise and power fluctuations in the excitation light source, but these can often be minimized, whereas shot noise is a fundamental limitation. Another simplification in this model is that, in practice, APs are usually over-sampled such that a single AP is spread out over several time bins. Based on the expected waveform of the AP, a matched filter can be constructed to extract APs optimally. This approach will yield improved performance for real APs, which are not square pulses, but it does not circumvent the fundamental effect of shot noise on detection efficiency.
| Signal Detection Model of Voltage Reporters |
|---|
|
|
|---|
FPM/FPM), 3) absolute brightness (FPM), 4) labeling specificity (FPM/FNS), and 5) cellular perturbation.
|
|
A simple version of this can be modeled by increasing W, the duration or width of the optical signal associated with the AP in the hypothetical case considered previously (FIGURE 2
). FIGURE 4B
shows that increasing the duration of the optical event dramatically increases detection efficiency. Increasing the duration of the optical signal is tantamount to enhancing the brightness of the indicator, since both increase the number of photons collected per AP, decreasing the effect of shot noise.
The response kinetics of fast VSDs are faster than the kinetics of APs themselves (12, 41). However, the response kinetics of two-component VSDs and hybrid GFP-dye systems are variable. High hydrophobicity of the mobile dye tends to correlate with fast kinetics (19), but hydrophobic dyes are harder to load into tissue (9). Nevertheless, some mobile dyes are sufficiently fast for AP recording. In the two-component VSD case, the use of highly hydrophobic versions of the mobile oxonol dye decreased the time constant to 0.4 ms, while in the hybrid GFP-dye case, the use of the hydrophobic quenching molecule dipicrylamine (DPA) yielded a time constant of 0.53 ms. In all cases, there is no mechanism for asymmetric kinetics, which makes APs more difficult to detect.
GFP-channel fusions are unique among voltage reporters in that the voltage-dependent conformational changes of voltage-gated cation channels exhibit hysteresis as a result of inactivation (2). In other words, a channel goes through a sequence of states on cell depolarization, but on repolarization the channel does not run through the same sequence in reverse but rather goes through a different set of states. This provides a potential biophysical substrate for the asymmetric on/off kinetics exhibited by the ideal voltage reporter. In practice, SPARC (3) and VSFP1 (38) show symmetric, sub-millisecond kinetics similar to VSDs. FlaSh, however, displays asymmetric responses to voltage pulses with an "on" time constant of ~23 ms and an "off" time constant of ~105 ms (43). Although this is too slow for AP recording, improved versions of FlaSh have "on" time constants as low as 9.7 ms (21), and it is possible that future versions of FlaSh will have even faster kinetics.
Voltage sensitivity (
FPM/FPM)
Voltage sensitivity is typically reported in terms of measured
F/F per 100 mV, determined under wide-field illumination in cultured cells, Xenopus oocytes, or squid axons. This measurement corresponds to
FPM/(FPM + FNS + B), which is necessarily an underestimate of
FPM/FPM, the true voltage sensitivity of the voltage reporter, as the "nonproductive" terms FNS and B also add to the denominator. Higher
FPM/FPM is always better, but the important point is that the
FPM/FPM necessary for efficient AP detection is highly dependent on the other parameters.
The voltage sensitivity of VSDs ranges up to 23%
F/F per 100 mV in cultured cells (35), although excitation of the red edge of the absorption spectrum of some dyes has been shown to increase this range to 70%
F/F per 100 mV (23) at the expense of a reduced total photon count. VSDs that give fast voltage-dependent SHG signals, such as FM4-64, yield lower sensitivity of up to 12%
F/F per 100 mV (37), but the SHG signal is generated selectively by membrane-bound dye molecules, decreasing FNS and increasing measured
F/F in intact tissue. However, it should be noted that the SHG signal is collected not by the objective, but by a detector placed on the opposite side of the specimen, limiting applicability to a subset of intact tissue preparations. Two component VSDs based on coumarin-oxonol FRET pairs yield high sensitivity (up to 80%
F/F per 100 mV) for slowly responding oxonols, but oxonols fast enough for AP recording yield lower sensitivities of ~8%
F/F per 100 mV (19). The EGFP-DPA FRET pair in the hVOS system exhibits a comparably larger
F/F of up to 34% per 100 mV (10).
In GFP-channel fusions, each fluorescent molecule senses membrane voltage with multiple gating charges. As a result, GFP-channel fusions are theoretically capable of the greatest voltage sensitivity. In practice, however, GFP-channel fusions exhibit some of the lowest voltage sensitivities, in the range of <5%
F/F per 100 mV. This is an area for great improvement in this class of voltage reporters, since wild-type GFP exhibits low conformational sensitivity due to the rigid ß-barrel structure shielding the chromophore (33). It is likely that the use of GFP variants with increased conformational sensitivity could greatly increase voltage sensitivity.
Absolute brightness (FPM)
The total photon flux a fluorophore can generate is one of the most important factors in practical imaging, and voltage reporters are no exception. There are two reasons for this. First, minimizing the relative contribution of background fluorescence B to total fluorescence increases the measured fractional fluorescence change
FPM/(FPM + FNS + B). More importantly, photon shot noise scales as the square root of total photon number, i.e., decreases in relative importance as the photon count increases. Optical recording of APs in intact tissue involves short temporal windows and single-cell resolution, which usually requires laser scanning microscopy. As a result, the number of photons per AP is already low, and voltage reporters generating low numbers of photons can make AP detection very inefficient, if not impossible.
Evaluating the example of AP detection with a false-positive rate PFP = 104 and an AP lasting 1 ms (FIGURE 4D
), we see that detection efficiency drops off precipitously below 3,000 photons per AP, or 3 x 106 photons per second. Although typical laser scanning microscopy experiments with some voltage reporters can generate >107 photons/s from a single cell, this has not been demonstrated with all voltage reporters. Despite great practical importance, there is no standard method for measuring or reporting brightness of voltage reporters. In theory, the brightness of a fluorophore could be quantified by taking into account the number of fluorophore molecules per cell, extinction coefficient (or two-photon absorption cross section), and quantum yield. In practice, however, fluorescence intensity is dependent on the power of the excitation light source, so photobleaching and phototoxicity become the limiting factors. With VSDs, this is usually the case.
One significant advantage of GFP-based voltage reporters is that GFP tends to have decreased photo-bleaching and phototoxicity relative to small molecule dyes. This is likely due to the ß-barrel structure of GFP, which shields the chromophore from molecular oxygen (33). In addition, VSDs are in direct contact with membrane lipids, leading to lipid peroxidation (18), whereas the GFP of genetically encoded voltage reporters usually protrudes into the juxtamembrane cytoplasmic space, where the cell is less vulnerable to oxidative stress.
However, brightness is a significant problem with GFP-channel fusions, since incorporation of GFP into the channel protein typically decreases GFP folding efficiency severely (4). High absolute brightness is among the advantages of the hVOS GFP-dye system (10). Unlike GFP-channel fusions, in which the GFP must fold within the topological constraints of an integral membrane protein and reach the plasma membrane via transport through the secretory pathway, the prenylated EGFP-F fluorophore in hVOS is expressed as a soluble protein and membrane-targeted via post-translational lipidation (10, 36). As a result, each cell tends to produce more properly folded GFP molecules, resulting in more favorable photon statistics.
Nonspecific labeling (FNS/FPM)
One of the primary differences between VSD performance in cultured cells and thick tissue specimens is due to nonspecific labeling. Nonspecific labeling arises from multiple sources, including VSD binding to nonexcitable cells and acellular tissue, or GFP molecules not localized to the plasma membrane. An ideal voltage reporter would have an FNS/FPM of zero, as any increase in FNS/FPM acts to decrease the effective
F/F. FIGURE 4D
shows that increases in FNS lead to a significant decrease in detection efficiency.
The greatest drawback of VSDs is limited labeling specificity. Since the dyes are nonselectively added to the tissue, nonneuronal cells and acellular tissue are also labeled, increasing background fluorescence dramatically. The use of SHG, rather than fluorescence, with VSDs largely eliminates the nonspecific background, as only membrane-associated VSD molecules generate an SHG signal.
GFP-based voltage reporters are often believed to have superior labeling specificity, but this is not always the case. Although expression is limited to a particular cell population, not all of the fluorescent GFP molecules are localized to the plasma membrane. GFP-channel fusions, for example, are translated in the rough ER and transported through the secretory pathway. Usually, significant amounts of GFP fluorescence are retained in the ER and Golgi (4). This may reflect quality control mechanisms preventing the export of channels that are misfolded due to GFP insertion, physiological control of channel densities on the cell surface, or both.
High membrane specificity is one of the great strengths of the hVOS system (10). Unlike GFP-channel fusions, the GFP is expressed as a soluble protein and posttranslationally targeted to the plasma membrane (10, 36). As a result, a significantly higher fraction of total GFP molecules are localized to the plasma membrane relative to GFP-channel fusions.
Cellular perturbation
Although not explicitly linked to the other parameters, the extent to which a voltage reporter perturbs normal cellular function also places limits on its utility. For example, VSDs are known to sensitize cells to photodamage, limiting the duration for which preparations can be illuminated and observed (12).
A more important, underappreciated perturbation is the increased capacitive load that results from adding mobile "probe" charges to sense the transmembrane potential. The additional capacitive load of fast VSDs is insignificant, but it is a problem in the context of two-component VSDs and hybrid GFP-dye systems (9, 10, 19, 20), where the movement of the mobile dye molecule has been shown to inhibit APs at high concentrations (10). The problem of capacitive load is most severe with GFP-channel fusions, as voltage-gated cation channels typically possess 1214 gating charges per channel (1, 42), leading to a steeper voltage dependence of charge movement and a higher capacitive load per mobile charge. The ratio of GFP molecules to mobile charges is also unfavorable from a signal detection standpoint. For example, if a GFP-channel fusion protein containing one GFP per channel tetramer is expressed at a level of 23,000 charges per µm2, which is typical for these experiments (7), then there are ~1,800 GFP molecules per µm2 of membrane. For a spherical cell of 20-µm diameter, the number of GFP molecules per cell is less than if the cells cytoplasm were filled with 1 µM GFP, the concentration at which GFP fluorescence equals cellular autofluorescence (29). Imaging a planar membrane containing 1,800 GFPs/µm2 with two photon microscopy, the GFP concentration inside the excitation volume (calculated as in Ref. 51) equals ~1 µM. Two-photon excitation of 1 µM GFP with 15 mW of laser power yields ~106 photons/s, which is a borderline value for detecting APs with a voltage sensor reporting APs with a 10% increase in fluorescence (FIGURE 4, C AND D
). Inefficient folding or maturation of GFP in the context of an ion channel fusion protein will, of course, further diminish the expected photon flux. Despite these low fluorescence levels, the addition of 23,000 mobile charges per µm2 increases the membrane capacitance by ~370%, which is far above the level shown to inhibit APs in vitro (10). As a result, capacitive load places severe restrictions on expression level and fluorescence yield of GFP-channel fusion proteins.
Although hVOS also increases membrane capacitance, the problem is less severe because GFP fluorescence quenching will occur even if there is less than one DPA molecule per GFP, since only a fraction of the GFP molecules are in the excited state at any given time. As a result, there is a much smaller capacitive load per GFP molecule, yielding a brighter signal with less perturbation to normal cellular function.
| Conclusion |
|---|
|
|
|---|
There is also significant room for improvement in voltage-sensitive reporter molecules, in particular those with a genetically encodable component. In designing such probes, it is wise to compare carefully the photon fluxes expected from a voltage reporter with those needed to detect an AP in a single trial. Such an analysis will take into account the brightness of individual molecules, the densities at which they can be inserted into the plasma membrane without disrupting normal physiology, the size of the membrane patch surveyed in a typical experiment, the voltage sensitivity and response kinetics of the probe, and the specificity of cell surface labeling. A VSD pioneer remarked that "in some fields one can be reasonably certain that [a given] path will bear fruit; if not in ones own laboratory, at least in other laboratories pursuing the same subject. . . . [But in other fields] it may be generations or centuries before it is known whether a chosen path was fruitful. The risk of devoting ones life to a dead end does lend excitement to the choice (11)." Keeping the principles of signal detection theory in mind should allow one to minimize this risk.
| Acknowledgments |
|---|
Work in the authors laboratory is supported by the National Institutes of Health, Office of Naval Research, and the W. M. Keck Foundation.